ID
Icons
Message
Cause/Action
129
Standard density above
warning limit
Check process conditions. Adjust the parameter value 'Up‐
per warning limit' to normal process conditions.
130
Standard density below
warning limit
Check process conditions. Adjust the parameter value
'Lower warning limit' to normal process conditions.
131
Standard density below
alarm limit
Check process conditions. Adjust the parameter value
'Lower alarm limit' to normal process conditions.
132
Standard volume flow
above alarm limit
Check process conditions. Adjust the parameter value 'Up‐
per alarm limit' to normal process conditions.
133
Standard volume flow
above warning limit
Check process conditions. Adjust the parameter value 'Up‐
per warning limit' to normal process conditions.
134
Standard volume flow be‐
low warning limit
Check process conditions. Adjust the parameter value
'Lower warning limit' to normal process conditions.
135
Standard volume flow be‐
low alarm limit
Check process conditions. Adjust the parameter value
'Lower alarm limit' to normal process conditions.
136
Totalizer 1 above alarm
limit
Check process conditions. Adjust the parameter value 'Up‐
per alarm limit' to normal process conditions.
137
Totalizer 1 above warning
limit
Check process conditions. Adjust the parameter value 'Up‐
per warning limit' to normal process conditions.
Diagnostics and troubleshooting
10.2 Fault codes and corrective actions
SITRANS FC330
180
Operating Instructions, 01/2019, A5E44030648-AB