ID
Icons Message
Cause/Action
62
Fluid temp. below limit
Increase the fluid temperature. If the problem persists, con‐
tact your local Siemens representative.
63
Fluid temp. above limit
Reduce the fluid temperature. If the problem persists, con‐
tact your local Siemens representative.
64
Frame temp. below limit
Increase fluid temperature and check that ambient temper‐
ature is within specified limits. If the problem persists, con‐
tact your local Siemens representative.
65
Frame temp. above limit
Reduce fluid temperature and check that ambient temper‐
ature is within specified limits. If the problem persists, con‐
tact your local Siemens representative.
66
Standard Deviation above
limit
Measurement continues with values from last successful
zero point adjustment. Improve conditions for automatic
zero point adjustment and repeat adjustment.
67
Zero Point Offset above
limit
Measurement continues with values from last successful
zero point adjustment. Improve conditions for automatic
zero point adjustment and repeat adjustment.
68
Zero point adjustment
failed
Measurement continues with values from last successful
zero point adjustment. Improve conditions for automatic
zero point adjustment and repeat adjustment.
69
Empty Tube Limit' excee‐
ded
Make sure that the sensor is filled with liquid and that the
liquid density is within the specified 'Empty Tube Limit'.
70
Too little fluid in tube
Make sure that the sensor is filled with liquid.
Diagnostics and troubleshooting
10.2 Fault codes and corrective actions
SITRANS FC330
174
Operating Instructions, 01/2019, A5E44030648-AB