7ML19981BC01
69
InterRanger
DPS300
Profile Records (P330 to P337)
The following parameters are for authorized Siemens service personnel or
Instrumentation Technicians familiar with Siemens echo processing techniques.
Use these features to record and save a total of up to 10 Echo profiles, initiated
manually (P330), or automatically (P331 et al). See Scope displays (P810) for
echo profile viewing hardware / software requirements.
If 10 Echo Profiles are already saved, addresses 1 through 10 are filled, the oldest
automatically initiated record is overwritten. Manually initiated records are not
automatically overwritten. All records are automatically deleted in the event of a
power interruption.
When a record is displayed, results are based on current programming (which
may have been altered since the record was saved). This permits the effect on the
echo profile to be observed when changing an echo parameter.
P330 Profile Record
In addition to serving as a library for profile records, this parameter provides two
functions:
•
manually records and saves echo profiles
•
displays an echo profile, recorded manually or automatically, e.g. via an
oscilloscope.
To select a record address:
Press
until address index is enabled.
Pressing
selects the desired address, 1 to 10, and displays the associated
parameter
Value:
`- - - -' = no record
` x# ' = record
Where:
X = A, automatically initiated
= U, manually initiated
# = transducer number
To manually record a profile:
Press
, and transducer* fires, and echo profile is recorded into the internal
scope buffer for display.
e.g. initial parameter display
e.g. address 2 selected, no record
saved