7ML19981BC01
47
InterRanger
DPS300
Potable Water Clarifiers
10 = Stationary mounting with bottom scrapers only.
11 = Stationary mounting without skimmers or scrapers.
12 = Rectangular clarifier with travelling bridge arrangement.
13 = Circular clarifier with rotating bridge arrangement.
Mining Clarifiers
14 = Stationary mounting with bottom scrapers only.
15 = Stationary mounting with surface skimmers and bottom scrapers.
Settling Tanks
16 = Stationary mounting without skimmers or bottom scrapers
(suspended solids >1%)
17 = Stationary mounting without skimmers or bottom scrapers
(suspended solids <1%)
P005 (G) Units
Enter the units of measure desired for programming Empty (P006) and Span
(P007).
Values:
1 = meters (m) (preset)
2 = centimeters (cm)
3 = millimeters (mm)
4 = feet (ft)
5 = inches (in)
P006 Empty
Enter the
maximum distance
(transducer face to clarifier bottom) to be measured,
in Units.
This value is preset to 4.000 m or equivalent Units (P005).
The value entered automatically sets Span (P007) to the maximum recommended
value.
Values
: 0.000 to 9999