ID
Icons
Message
Cause/Action
57
Malfunction in sensor driv‐
er
Contact your local Siemens representative.
58
Unstable driver oscillation Contact your local Siemens representative.
59
Mass flow out of specifica‐
tion
Reduce the flow. If the problem persists, contact your local
Siemens representative.
60
Volume flow out of speci‐
fication
Reduce the flow. If the problem persists, contact your local
Siemens representative.
61
Density out of specifica‐
tion
Contact your local Siemens representative.
62
Fluid temp. below limit
Increase the fluid temperature. If the problem persists, con‐
tact your local Siemens representative.
63
Fluid temp. above limit
Reduce the fluid temperature. If the problem persists, con‐
tact your local Siemens representative.
64
Frame temp. below limit
Increase fluid temperature and check that ambient temper‐
ature is within specified limits. If the problem persists, con‐
tact your local Siemens representative.
65
Frame temp. above limit
Reduce fluid temperature and check that ambient temper‐
ature is within specified limits. If the problem persists, con‐
tact your local Siemens representative.
66
Standard Deviation above
limit
Measurement continues with values from last successful
zero point adjustment. Improve conditions for automatic
zero point adjustment and repeat adjustment.
67
Zero Point Offset above
limit
Measurement continues with values from last successful
zero point adjustment. Improve conditions for automatic
zero point adjustment and repeat adjustment.
68
Zero point adjustment
failed
Measurement continues with values from last successful
zero point adjustment. Improve conditions for automatic
zero point adjustment and repeat adjustment.
69
Empty Tube Limit' excee‐
ded
Make sure that the sensor is filled with liquid and that the
liquid density is within the specified 'Empty Tube Limit'.
70
Too little fluid in tube
Make sure that the sensor is filled with liquid.
71
Parameter storage mal‐
function
Turn off the power, wait 5 seconds and turn on the power
again. If the problem persists, contact your local Siemens
representative.
72
Internal error in sensor
Contact your local Siemens representative.
73
Internal error in sensor
Contact your local Siemens representative.
74
Internal error in sensor
Contact your local Siemens representative.
75
Internal error in sensor
Contact your local Siemens representative
76
Internal error in sensor
Contact your local Siemens representative.
Diagnostics and troubleshooting FCT010
10.2 Fault codes and corrective actions
SITRANS MASS 2100 & FC300 (FCT010/DSL)
Operating Instructions, 06/2020, A5E39788682-AB