Table 11-3
Mass flow accuracy
Sensor size
DI 1.0
DI 1.5
DI 2.1
DI 3
DN4 HP
DN4 C22 DN4 316 DI 6
DI 15
Qmin - minimum
flowrate
[kg/h] (lb/min)
On
request
0.1
(0.22)
On
request
1
(2.2)
On
request
1
(2.2)
1
(2.2)
5
(11.0)
20
(44.1)
Qnom - nominal
flowrate
[kg/h] (lb/min)
9,8
(21.6)
19.5
(43)
42
(92.6)
90
(198.4)
35
(77.2)
95
(209.4)
140
(308.6)
500
(1102)
3800
(8378)
Qmax - maximum
flowrate [kg/h] (lb/
min)
15
(33)
30
(66)
50
(110)
250
(550)
40
(88)
255
(562)
350
(770)
1000
(2200)
5600
(12350)
Max. zero point sta‐
bility
[kg/h]
On
request
0.001
(0.002)
0.001
(0.002)
0.01
(0.02)
On
request
0.01
(0.02)
0.01
(0.02)
0.05
(0.11)
0.2
(0,4)
Measuring accuracy
[%]
0.3%
0.1%
0.1%
0.1%
0.3%
0.1%
0.1%
0.1%
0.1%
Repeatability error
[%]
0.05%
0.05%
0.05%
0.05%
0.05%
0.05%
0.05%
0.05%
0.05%
Table 11-4
Density accuracy (liquids)
Sensor size
DI 1.0
DI 1.5
DI 2.1
DI 3
DN4 HP
(DI 2.0)
DN4 C22
(DI 3.0)
DN4 316
(DI 3.5)
DI 6
DI 15
Density accuracy,
standard calibration
[g/cm³]
n/a
0.005
0.005
0.005
n/a
0.0025
0.007
0.005
0.005
Density accuracy, ex‐
tended calibration
[g/cm³]
n/a
0.001
0.001
0.0015
n/a
0.0025
0.007
0.0015
0.0005
Density repeatability
[g/cm³]
n/a
0.0002
0.0002
0.0002
n/a
0.0002
0.0002
0.0002
0.0001
Table 11-5
Media temperature accuracy
Sensor size
DI 1.0
DI 1.5
DI 2.1
DI 3
DN4 HP
(DI 2.0)
DN4 C22
(DI 3.0)
DN4 316
(DI 3.5)
DI 6
DI 15
Media temperature
accuracy [°C]
±0.5
±0.5
±0.5
±0.5
±0.5
±0.5
±0.5
±0.5
±0.5
Media temperature
repeatability [°C]
±0.25
±0.25
±0.25
±0.25
±0.25
±0.25
±0.25
±0.25
±0.25
Brix error
[°Brix]
n/a
0.2
0.2
0.3
n/a
0.3
0.3
0.3
0.1
Technical data
11.2 Performance
SITRANS MASS 2100 & FC300 (FCT010/DSL)