Table C-16
Volume flow calibration
Modbus ad‐
dress
Data type /
Size [bytes]
Parameter
Description
Default value
[unit]
Value
range
Access
level
2103
Float / 4
Maximum Volume
flow Capacity
Maximum volume flow
measurement capacity of
the sensor
Sensor size
specific [m
3
/
s]
1)
0 to 0.177
Read only
1)
: See Sensor dimension dependent default settings (Page 140).
Table C-17
Mass flow calibration
Modbus ad‐
dress
Data type /
Size [bytes]
Parameter
Description
Default value
[unit]
Value
range
Access
level
2101
Float / 4
Maximum Mass flow
Capacity
Maximum mass flow
measurement capacity of
the sensor
Sensor size
specific [kg/
s]
1)
0 to 1023
Read only
2402
Float / 4
Calibration Factor
Factory-set sensor-specific
calibration factor.
The calibration factor is
shown on the sensor name‐
plate
-
Min: 5.00E
+07
Max: 4.29E
+09
Read only
1)
: See Sensor dimension dependent default settings (Page 140).
Table C-18
Density calibration
Modbus ad‐
dress
Data type /
Size [bytes]
Parameter
Description
Default value
[unit]
Value
range
Access
level
2428
Float / 4
Density Calibration
Offset
Specify an offset in the den‐
sity flow calculation
-
-14 000 to
+14 000
Read only
2430
Float / 4
Density Calibration
Factor
Specify gain factor in the
density flow calculation
-
-21474835
83 to
214748358
3
Read only
2432
Float / 4
Dens. Comp. Tube
Temp.
Specifies a tube temperature
coefficient in the density cal‐
culation
-
-0.001953
to
+0.001953
Read only
2434
Float / 4
Dens. Comp. Frame
Temp.
Specifies a frame tempera‐
ture coefficient in the density
calculation
-
-0.001953
to
+0.001953
Read only
Modbus communication FCT010
C.4 Modbus holding registers tables
SITRANS MASS 2100 & FC300 (FCT010/DSL)
172
Operating Instructions, 06/2020, A5E39788682-AB