Table C-5
Mass flow
Modbus ad‐
dress
Data type /
Size [bytes]
Parameter
Description
Default value
[unit]
Value
range
Access
level
2125
Float / 4
Low Mass flow Cut-
Off
Set mass flow limit for low
flow cut-off.
Below this limit mass flow
output is forced to zero.
If Low Flow Cut-Off is set to 0,
the cut-off functionality is
disabled.
Notice:
It is recommended to set a
lower value for gas applica‐
tions.
Sensor size
specific [kg/
s]
1)
0 to 1023
Read /
Write
2426
Float / 4
Mass flow Correction
Factor
Specify correction factor for
use in the mass flow calcula‐
tion
1
-1.999 to
+1.999
Read /
Write
1)
: See Sensor dimension dependent default settings (Page 140)
Table C-6
Volume flow
Modbus ad‐
dress
Data type /
Size [bytes]
Parameter
Description
Default value
[unit]
Value
range
Access
level
2170
Float / 4
Low Volume flow Cut
Off
Define the numerical volume
flow value below which the
volume flow output is forced
to zero.
Sensor size
specific [m
3
/
s]
1)
0 to 0.177
Read /
Write
1)
: See Sensor dimension dependent default settings (Page 140)
Table C-7
Density
Modbus ad‐
dress
Data type /
Size [bytes]
Parameter
Description
Default value
[unit]
Value
range
Access
level
2127
Float / 4
Empty Tube Limit
Define threshold value of
empty tube
500 [kg/m
3
]
-14 000 to
+14 000
Read /
Write
2129
Unsigned / 2 Empty Tube Detection Set automatic detection of
Empty Tube On/Off
0 = off (Empty tube is off).
1 = on (a density value below
Empty Tube Limit triggers an
alarm. All flow rate values
are forced to zero %).
0
0 to 1
Read /
Write
Modbus communication FCT010
C.4 Modbus holding registers tables
SITRANS MASS 2100 & FC300 (FCT010/DSL)
164