Automatic zero point adjustment
The device measures and calculates the correct zero point automatically.
The automatic zero point adjustment of the flowmeter is set by the following parameters:
• Duration (Modbus address 2135)
• Start Zero Point Adjustment (Modbus address 2180)
When zero adjust is initiated by selecting Start Zero Point Adjustment, the mass flow values are
acquired and totalized for the configured period (Duration). The default zero point adjustment
period (30 s.) is normally sufficient for a stable zero point measurement.
Note
Extremely low flow quantity
If the flow quantity is extremely small, extremely precise measurement is necessary. In this case,
a long zero point adjustment period can be selected for improved zero point adjustment.
Zero point calculation
During zero point adjustment, an average value is automatically calculated using the following
formula:
Zero point offset value
Average of N flow values
N
x
x
N
i
i
∑
=
≡
1
x
i
is an instantaneous flow value sampled in
the time domain
N = Number of samples during zero point ad‐
justment
The offset value must be within the determined Zero Point Offset Limit (Modbus address 2140).
Note
Exceeded zero point offset limit
If the offset value is greater than the configured limit, proceed as follows:
• Check that the tube is completely filled and that the flowrate is absolute zero.
• Check the validity of the configured zero point offset limit.
• Repeat the zero point adjustment.
Parameter assignment FCT010
8.1 Functions
SITRANS MASS 2100 & FC300 (FCT010/DSL)
Operating Instructions, 06/2020, A5E39788682-AB
77