LABOKLAV Eco 80 / 135
User Manual
20
break shall happen. Normally the program is going into the pressure reduce phase for solid goods
or into the cooling phase for liquids. The program is only using the parameters for standard
programs of this sterilization good! Especially if liquids were chosen, the program break cause a
very long waiting time until unlocking the lid! In or after the cooling phase of liquid programs a
program cancellation is not possible!
In case of a liquid program the activated thermo lock can not disabled by using the program
cancellation function!
In case of a program cancellation before regular finish of
sterilization phase the sterilization goods are not sterile! The
sterilization needs to be repeated!
Closing the lid
Press the lid with your left hand down against the locking spindle. Push the “down button” with right
hand and keep it pressed until the locking procedure has completely finished. The locking
procedure has 3 steps: running first turn, sopping for 3 seconds, running 2
nd
turn. Breaking the
procedure by let loose the button an error message for not correctly closed lid will be generated!
You have to open again and repeat to close again.
Open the lid
If the chamber is closed and you want to open please press “Up button” once. Lid will be
automatically opened. After program has finished and you quit the program, the autoclave is
automatically opening the lid. Do not get frightened if there is sometimes a loud noise (bang) during
lid opening process. This noise is normal.
Draining the air and steam
To remove the air and steam the unit is equipped with an outlet to connect to a drainage system. If
the unit is connected with cold water supply the program is automatically protecting the condensate
drain against over heating by direct draining with hot steam. If it is not possible to connect the unit
with central drainage please use a condensate collector tank. If you do so please make sure the
cooling water is not connected and is closed with a cap.
Documentation of sterilization cycle
The device is prepared for 2 different kinds of cycle documentation. The first is the connection to a
standard IBM compatible PC via RS485 interface in the back of the unit. For read out of data a
special interface transformer is necessary (RS 485 to USB adapter). Additionally you need to install
documentation software DOKUMENTATOR
The second possibility is building in a standard needle printer for 40 characters per line with serial
interface (CBM 910). The sterilizer has a built in memory with up to 4 MB and is automatically
storing the data of each cycle until memory is full. Then start program over writing the oldest data.
Data readout by one of the 2 possibilities is possible later on.
The parallel use of the 2 documentation possibilities is available under special conditions too.