LABOKLAV Eco 80 / 135
User Manual
24
Er0109- Door closing function: locking cycle was broken
message:
Door locking cycle was broken. Please open again
description:
error occurs if locking operation was broken by the operator, it needs new
opening operation and then closing again, or the unit can not see closed
GS05 (motor position switch) or GS01 (door position switch) Maybe K1 is not
correctly connected)
Er0201- Archiv Data Error:
message:
Archive data Error
description:
error occurs if new software was installed with reorganization of archive
structure: delete the archive data (Service technician)
8. Maintenance
The sterilizer should get regular cleaning, maintenance and service. Some parts are regular to be
changed completely to protect the device against damage or mistakes in sterilization cycle. The
simple cleaning and maintenance activities can be done by the operator without problems.
Special services can be done by specially trained service stuff only! All inspection activities acc. to
pressure vessel regulations / local regulations for pressure vessels and electrical installations need
special trained service stuff! We recommend to order one regular safety inspection per year and to
connect this with a regular maintenance for the vessel, pressure parts and electrical installations.
Your distributer is authorized to tell recommend a trained service partner.
We recommend using a device book that helps to document
all cleaning, maintenance and service repair activities at the
device.
Maintenance or repair activities that need to open the housing
are not allowed to be done by untrained personnel stuff!
For maintenance or repair activities that need to open the
housing the electrical power supply must be disconnected!
Inside the housing dangerous electrical voltage can kill or
hurt!
After work at electrical installations some electrical test are
necessary to do. Please respect the local rules and
regulations!