LABOKLAV Eco 80 / 135
User Manual
18
The display shows program no. P1. Here the program type is shown; if special program name
was given it will replace the program type. The main program parameters are shown for fast
identification of the program cycle. In this case you will find the deaeration and the relevant
sterilizing parameters temperature and time. The most important sensor data of chamber
temperature, chamber pressure and, if configured, reference sensor are displayed. Date and
time are always indicated for documentation issues.
The program P1 was started but lid was not completely
closed. So the device generates an error message. The
error message can be cleared by pressing the Escape
button if the cause of the error was cleared.
The program was started; the program has initialized the
parameters regular and is now in deaeration phase step
1.
Deaeration has finished, the program is now in heating
phase. Setpoint for finish of this phase is reaching the
134°C chamber temperature.
The sterilizer is in sterilization phase, 2 minutes before