is positioned should be from the same size, form and filled with same volume of the largest
single volume of the sterilization goods.
- Protection against steam out coming from chamber – Steam production is switched off
immediately if the lid is opened.
- Protection against over heating the steam generator – The chamber is protected against
over heating over temperature switch. The switch is self resetting when temperature is low
again. While the temperature is high it is not possible to open the chamber! You have to wait
until chamber cools down by them selves. Over temperature switch is breaking the program!
11. Definition of feed water quality
Acc. to EN 285 – “Steam sterilizers”, app. B / EN 13060 – small size steam sterilizers App. C
Feed water
Condensate
Residual dry matter
≤
10 mg/l
≤
1.0 mg/kg
Silica oxide, SiO
2
≤
1 mg/l
≤
0.1 mg/kg
Iron
≤
0.2 mg/l
≤
0.1 mg/kg
Cadmium
≤
0.005 mg/l
≤
0.005 mg/kg
Lead
≤
0.05 mg/l
≤
0.05 mg/kg
Other heavy metals, except for iron,
cadmium, lead
≤
0.1 mg/l
≤
0.1 mg/kg
Chlorines
≤
2 mg/l
≤
0.1 mg/kg
Phosphates
≤
0.5 mg/l
≤
0.1 mg/kg
Conductivity (at 20
°
C)
≤
15
µ
S/cm
≤
3
µ
S/cm
pH
5 do 7
5 do 7
Colour
Colourless, clean, no deposit
Colourless, clean, no deposit
Hardness
≤
0.02 mmol/l
≤
0.02 mmol/l
NOTE 1: Using water of contamination greater than specified above for steam generation can
considerably reduce the sterilizer life and void the manufacturer’s warranty.
NOTE 2: The condensate should be derived out of the steam collected during sterilizing cycle with the
chamber empty.
Tests for conformance are performed with commonly used analytic methods.