LABOKLAV Eco 80 / 135
User Manual
23
Er0053- Chamber pressure under available range
message:
Chamber pressure is under available range
description:
error occurs if readout from PIS02 sensor is less than admissible minimum
(Service check measuring channels, CH3: Pk)
Er0054- Chamber pressure over available range
message:
Chamber pressure is over available range
description:
error occurs if readout from PIS02 sensor is more than admissible maximum
(Service check measuring channels, CH3: Pk)
Er0055- Reference temperature under available range
message:
Reference temperature is under available range
description:
error occurs if readout from TIC22 sensor is less than admissible minimum
(Service check measuring channels, CH2: Tref)
Er0056- Reference temperature over available range
message:
Reference temperature is over available range
description:
error occurs if readout from TIC22 sensor is more than admissible maximum
(Service check measuring channels, CH2: Tref)
Er0101- Door of sterilizer opened at cycle start
message:
Door of sterilizer is open (GS01)
description:
error occurs if controller detects that GS01 switch is not closed
on cycle
startup (see locking symbol in the display)
Er0106- Exceeded maximum available cycles number for filter counter
message:
Filter counter: exceeded cycles number
description:
error occurs if counted filter cycles number is higher than admissible
maximum – the filter elements needs to be changed and filter counter needs
RESET (code 0911/Mainmenu/Statistical Data)
Er0108- Pressure in chamber outside atmospheric pressure range
message:
Pressure in chamber not correct (PLS02)
description:
error occurs if chamber pressure is above 0,15 bar (relative) or zero pressure
switch (PLS02) is disconnected (chamber pressure is outside atmospheric
pressure range) and operator tries to open a door, check K11 if disconnected
error is generated