LABOKLAV Eco 80 / 135
User Manual
27
9. User replaceable accessories and spare parts
Spare part
Drawing number
Article number
Lid seal Laboklav ECO 80
Laboklav 80 ECO
50-05-10000-003
Lid deal Laboklav ECO 135
Laboklav 135 ECO
50-05-10000-005
Magnetic valve Laboklav ECO
Laboklav 80 + 135 ECO
50-05-10000-014
Filter 0,3
µ
m / 99,5%
Laboklav 80 + 135 ECO
40-0719-142-014
Paper rolls for printer
Laboklav 80 + 135 ECO
50-05-30001-195
Carbon stripe for Printer
Laboklav 80 + 135 ECO
50-05-30001-183
Printer for Laboklav ECO
Laboklav 80 + 135 ECO
20.0000.000.300
Basket Ø 39,5cm x 30cm
Laboklav 80 ECO
30.0400.000.300
Bucket Ø
Laboklav 80 ECO
30.1400.000.300
Basket Ø
Laboklav 135 ECO
30.0500.000.300
Bucket Ø
Laboklav 135 ECO
30.1500.000.300
Basket with closed bottom Ø 39,5cm x 25cm Laboklav 80 ECO
30.2400.000.250
Basket with closed bottom Ø
Laboklav 135 ECO
30.2500.000.250
Condensate collection canister
Laboklav 80 + 135 ECO
20.0000.000.500
10. Description of safety devices
The steam sterilizer is equipped with different safety devices. The safety devices protect the user
against injury and safe the sterilization process. Mechanical and electronic safety devices are built
in and realize in sum a safety concept with different safety functions.
- Protection against over pressure – If control board is measuring a chamber pressure of
more than 345 kPa absolute pressure (2.45 bar relative pressure) an alarm is generated
(including error message) and the heating function is switched of and the unit breaks the
program. With 2.8 bar relative pressure the safety valve is opening and chamber pressure is
reduced mechanically! The steam is blowing into the housing contact with the steam is not very
dangerous because it is saturated after blowing out.
Attention: To check the safety valve a special program can be implemented that is bridging
the safety functions of over pressure protection. This program will be implemented on special
order by the customer only! Blowing off the safety valve in the condition that the housing I not
opened can damage the electronic board!
- Protection against opening the chamber while over pressure is inside – The device has
a built in thermo locking function. The device opens the thermo lock when pressure is low only.
The pressure is checked by pressure sensor and an additional pressure switch that detects
normal pressure. The opening mechanism is calculated to open when chamber pressure is low.
These 3 safety functions give good protection against opening while pressure in chamber is
high.
- Protection against opening the chamber while temperature of liquids is too high – One
part of the thermo locking system is the measurement of the temperature inside liquids by the
reference sensor. The device is unlocking the lid in liquid programs when temperature is lower
than programmed removing temperature is reached only. The flask where the reference sensor