7. Troubleshooting
Er0001- Door of sterilizer opened during a cycle
message:
Door of sterilizer is open during a cycle
description:
error occurs if controller detects that cover is not closed (GS01 switch is not
closed or K1 has no contact up from year 2008) during a cycle,
Er0004- Exceeded maximum available pressure in chamber
message:
Exceeded max. available pressure in chamber
description:
error occurs if pressure in chamber is higher than 245,0kPa (relative scale):
problems with deaeration / pressure sensor
Er0006- Exceeded maximum available temperature in chamber
message:
Exceeded max. available temperature in chamber
description:
error occurs if temperature in chamber is higher than 145,0 ºC
Er0040- Disconnected steam generator thermal switch
message:
Disconnected steam gen. thermal switch (THSZ01)
description:
error occurs if switch THSZ01 is disconnected,
Er0041- Disconnected chamber thermal switch
message:
Disconnected chamber thermal switch (THZ11)
description:
error occurs if switch THZ11 is opened and switch THSZ01 is closed,
Er0042- Disconnected safety line
message:
Disconnected safety line
description:
error occurs if switch THZ11 or switch THSZ01 is disconnected,
Er0051- Chamber temperature under available range
message:
Chamber temperature is under available range
description:
error occurs if readout from TIC21 sensor is less than admissible minimum
(Service check measuring channels, CH2: Tk)
Er0052- Chamber temperature over available range
message:
Chamber temperature is over available range
description:
error occurs if readout from TIC21 sensor is more than admissible maximum
(Service check measuring channels, CH2: Tk)