LABOKLAV Eco 80 / 135
User Manual
5
1. List of delivered parts
Device in ordered specification (options)
Connecting pipes for water supply steam / condensate remove
Bottom sheet
Documents including user manual, pressure vessel papers (conformity declaration), safety valve
calculation, warranty declaration
Baskets acc. to order
2. Unpacking
Unpacking, installation, instructing into the use of autoclave has to
be done by certified / authorized stuff only. Lokal regulations acc.
to PED 23/97/EU may regulate the installation process.
While unpacking and installation the technician should have attention to the feet and rollers of the
autoclave to protect the unit against damage! The feet in the front are used to fix the autoclave in a
straight position.
3. Installation
Preparation of the unit
The floor in the room should be waterproof. The sterilizer require safe foundations, and it should be
located in close distance to a floor drain, to allow draining of condensate and water. The room
dimensions should ensure comfort in operation, while ensuring the minimal distances (min. 10 cm
to the wall and next device in room).
Due to large amount of heat generated during the sterilizer operation, the room in which it is seated
should be fitted with a mechanical intake and exhaust ventilation system, providing 6÷10
exchanges of air per hour. The ventilating hood should be installed above the sterilizer.
By using the turnable leeds the chamber of the device should be adjusted to remove the
condensate from chaber easely.
Power supply
The steam sterilizer is equipped with a 2, 5 m long net supply cable. The device is configured to be
connected to an electrical system CEE 3P+N+GND with a voltage of 400V AC, 50 Hz, 16 A
(Laboklav ECO135) or 1N 230V 16A (Laboklav ECO 80). For commercial use we recommend to
use an additional fault current protection switch. For fast switch off a central main power switch
should be installed (see page 6).
If the device is connected to a power supply with wrong or
without correct ground connection it may endanger the
operator of the device by dangerous electrical voltage.
Connecting the device to a power supply with lower capacity
than 1x/3x16 A may cause an over load or heating up the