LABOKLAV Eco 80 / 135
User Manual
14
Switching on
After regular installation and connecting to media / power supply the device is ready for use.
Standing in front of the device you will find the main switch at the right side of the key pad.
Switching on the main switch the display shows the software version short time. When switched on
the device is ready for use.
Do not manipulate the device! If the device is not going on
please check the main breakers of the unit (4 – chap.1) and
power supply. If no reason can be found please inform a
trained service stuff to check the device internally.
Loading the device
We suggest using the standard baskets and drums we optionally offer for the special use in
Laboklav ECO steam sterilizers.
While loading and unloading the chamber, please consider
the chamber and surface of the device can be hot! Even so
the loading goods and loading system parts like baskets!
While loading and unloading you are in danger of burning!
Use adequate clothes to be protected against burning like
temperature stable isolating gloves etc.
Check the water before putting baskets inside the chamber. Fill demineralized feed water
up to the height of the bottom sheet inside the chamber manually.
Preparation of sterilization goods
Solid sterilization goods should be cleaned before sterilizing. Cleaning is a basic part of sterilization
process. It reduces the numbers of microorganism on the surface. The sterilization is not a cleaning
process! Microorganism will be killed or deactivated but the rest of material that the microorganism
consists of will be on the surface after the sterilization has finished! Often these particles act as
pyrogenes!
Solid sterilization goods should be cleaned before
sterilization! Only this gives warranty of a high quality
sterilization cycle!
Wrapped instruments can not be handled by gravitation deaeration process. The sterilization would
be not validable. Waste bags should be opened. This is necessary to make sure the steam comes
in direct contact with sterilization good. If there are doubts that the sterilization process runs without
complications it should be validated!