31
6. OPERATION
6.1 Operation
Once the power is turned on, the display on the front of the door lights. The instrument will be ready to
run (user mode) in about 5 seconds, including the initial screen. If the REC indicator is unlit, press the
“REC” key to start recording operation.
6.1.1 Status after Initial Screen
(1) Display:
The “REC” indicator retains its status prior to power-off.
(2) Printout data:
All printing data and analog recording data are cleared after the power has
been switched off.
(3) Alarm and Diagnosis: Alarm indications and output information are not recoverable after the
power has been switched off.
(4) Data and Channel number are displayed as the following Fig. 6.1.1.
Fig. 6.1.1 Screen at Power-on
[Note]
(1) If electric power failure occurs, the initialization is performed after power recovery.
The recorder reverts to the initial status above.
(2) The printout data is eliminated in initialization. Printout operation does not continue after power
recovery when the power is turned off during printout.
(3) When the power is turned on, the display indicates “ ” while reading data in Fig. 6.1.1.
The number of indicated digits differs depending on the setting.
CAUTION
(1) This recorder has no power switch.
(2) Prior to switching the power on, make sure that the chart paper is set in the chart holder.
If the printer is activated with no chart paper set, the sprocket drum (cylindrical part) of the chart
holder may be damaged.
WARNING
When calibrating data, connect the equipment according to Section 9.5 of the instruction manual.
Data calibration must be performed correctly.
Operators possessing knowledge of electricity and control related matters should perform the
calibration after careful reading of Sections 7.2 and 9.5 of the instruction manual.
(Shinko Technos Co., Ltd. is not liable for damages incurred if calibration is performed other than
prescribed in the instruction manual.)
Returns to the mode before
power-OFF.
WARNING
Prior to switching the power on, make sure that the supply voltage meets the specifications for
the instrument, and the instrument is properly grounded.
REC ALM