100
9.3 Replacing Consumables
Replace consumables according to the following list.
No.
Name
Type
Period
Remarks
Quantity
1
Chart paper
H-10100
33 days
When chart feed speed is
20 mm/h
1
2
Ribbon cassette WPSR188A000001A
3 months
When chart feed speed is
20 mm/h
1
9.4 Adjusting Dot Printing Position (Point Calibration)
The following describes how to adjust printing position on the chart paper.
It is advised to adjust recording position annually for accurate recording.
How to Adjust
Leave the recorder as it is, and adjust it according to the instructions in Section 7.2.11 Point
Calibration. (p. 81, 82)
9.5 Calibration (Data Calibration)
Calibrate a process variable input every year in order to keep measurement accurate.
Devices required for calibration
Precision voltage (current) generator (Jointly use a digital voltmeter with accuracy of
0.02% or less as required)
Precision dial resistor
[Note]
(1) Over time, ribbon colors will begin to fade. To ensure clear recording, replace the ribbon
cassette as soon as possible.
(2) Use the ribbon cassette within one year after its purchase. Over time, ribbon colors will
begin to fade.
[Note]
Prior to calibration, allow warm-up for at least 30 minutes after switching the power ON.