4.4
Setting Numeric Values
In the UP Series, various settings such as external sensitivity calibration weight value input, checkweighing
for setting threshold values, and medium density during specific gravity measurement, are input as numerical
values in some cases. (see
8.2
,
8.3
,
11.1
,
11.5
,
11.7
for detail of each item.)
In a menu used to set numeric values, [MENU] and [ ] are both prompted and the digit to be input blinks.
Press the [PRINT] key to move the blinking digit one place to the right.
Press the [UNIT] key to increase the value of the blinking digit by one.
Press the [O/T] key to store the displayed value in the balance memory.
…“SEt” is displayed when the value has been successfully saved.
…“Err” is displayed when the balance failed to save the value.
Press the [POWER] key to stop numeric entry. “Abort” is displayed briefly and the
display returns to the menu, one level up.
Notes
When setting decimal points (solid specific gravity measurement, liquid
specific gravity measurement, user unit multiplier settings)
1. Press the [PRINT] key repeatedly until the last digit is blinking.
The
▼
symbol or current decimal point will blink, and decimal point
position setting mode will start.
2. Press the [UNIT] key.
The selected decimal point position will move to the right (equivalent to a
value of × 10), so continue pressing until the desired position is blinking.
3. Press the [O/T] key to set the decimal point position. “SEt” is displayed
briefly to indicate that the setting is completed.
Use the optional AKB-301 Application Keyboard to easily set numerical values and decimals.
…