88
12
. Connections and Communications with External Equipment
12.5.3
Format
Set the balance output data format.
The standard format for the Shimadzu electronic balance:
Select menu item
77
[F-Fmt 1]
The old output format for the Shimadzu electronic balance:
Select menu item
78
[F-PrEEb]
The old output format is employed in the following models.
EB-500, 5000, 280, 2800, AEL-160, EB-50K (except -15)
Note
In this format, the number of the lowest place of menu item [70] [SC- id] is
assigned to identify the balance.
Compatible format for the PR and SR Series of Mettler-Toledo electronic balances:
Select menu item
79
[F-Fmt 3]
Available commands, functions and responses are limited
Compatible format for the IS Series of Sartorius electronic balances:
Select menu item
80
[F-Fms 4]
Available commands, functions and responses are limited
12.5.4
Communication Speed
Select the communication speed (300, 600, 1200, 2400, 4800, 9600, 19200, 38400BPS)
Number of “b-xxx” shows bps (bits/second). Baud rate and bps are the same value.
Select one of the menu items
81
~
88
12.5.5
Parity / Bit Length
Select the parity and bit length.
No parity, 8-bit length:
Select menu item
89
[P-no]
Odd number parity, 7-bit length:
Select menu item
90
[P-odd]
Even number parity, 7-bit length:
Select menu item
91
[P-EvEn]