![Shimadzu UP1023X Instruction Manual Download Page 52](http://html1.mh-extra.com/html/shimadzu/up1023x/up1023x_instruction-manual_1256959052.webp)
37
7
. Calibration
2
4
1
3
5
6
7
11
13
12
8
9
10
14
7.3
Calibration Setting
7.3.1
Selecting the Calibration Type
Set the calibration type that will be used in Calibration Execution.
To set up “Span calibration using the built-in weight”, (UPX Series only) Select menu item
1
.
To set up “Calibration check using the built-in weight”, (UPX Series only) Select menu item
2
.
To set up “Span calibration using external weights”, Select menu item
3
.
To set up “Calibration check using external weights”, Select menu item
4
.
Not applicable to a verified balance as a legal measuring instrument in using region
7.3.2
PSC Fully-automatic Calibration (UPX Series only)
The UPX Series is equipped with a function (PSC Function) which automatically carries out span
calibration when the ambient temperature changes so that such changes do not impact measurement.
This function can be turned ON and OFF using the following procedure.
1
To turn ON the PSC function, Select menu item
5
2
To turn OFF the PSC function, Select menu item
6
Summary of Contents for UP1023X
Page 3: ...No text...
Page 124: ...109 2 4 1 3 5 6 7 11 13 12 8 9 10 14 Appendices...