![Shimadzu UP1023X Instruction Manual Download Page 105](http://html1.mh-extra.com/html/shimadzu/up1023x/up1023x_instruction-manual_1256959105.webp)
90
13.1
Maintenance
If dirty
Use a soft damp cloth containing a neutral detergent to clean the balance.
The pan can be removed and washed with water. Verify that the pan is completely dry before replacing it
on the balance.
Avoid using organic solvents, chemicals, or dusting sprays as they may damage the coatings of the
balance or the display panel.
Attach the protective in-use cover (standard accessory) when the balance is used in an environment
where it is susceptible to being soiled.
13.2
Moving the Balance
To carry the balance
… hold the balance firmly with both hands.
To transport the balance
…use the shipping carton used to deliver the balance.
For the UPX Series, after disconnecting all accessories from the balance main body, rotate the
transportation screws on the bottom of the balance clockwise until they stop before placing the balance in
the packaging box.
The UPY Series is not equipped with transportation screws. Place the UPY Series balance in the
packaging box as-is to move.
!
Caution
When tightening the UPX Series transportation screws, turn the balance
main body over as described in steps
1
and
2
of "2.3 Installation". Make
absolutely sure to remove the pan supporter caps first.
Note
Using a verified balance as a legal measuring instrument in using
region:
Span calibration must be newly performed with the built-in calibration weight
after the balance has been moved and re-installed.
13.