74
11
. Application Functions
Notes
• Set the Stability and Response to Pouring mode (menu item
23
, See
18.2
) if faster response is required.
• Formulation Mode setting can be cleared without menu operation, by
pressing the [POWER] key for about three seconds.
• Cannot be used together with Pretaring (
10.7
).
• See
8.3
for details of stability detection.
• When formulation mode is ON, unit switching and minimum display
1d/10d switching are not allowed.
• When Formulation mode is ON, fully-automatic span calibration by PSC
(
7.3.2
) or Clock-CAL (
7.3.3
) is not performed. Weight symbol keeps
blinking when span calibration is necessary. Span calibration with the
built-in calibration weight or external weights can be performed between
sets of measurements (Formulation stand-by).
• Not more than one of these seven functions can be set at a time. Setting
the second one with the menu will automatically clear the previously set
function; Auto Print(
10.3
), Auto Zero(
10.4
), Peak Hold(
11.3
), Interval
Timer(
11.4
), Add-on mode(
11.5
), Animal Weighing mode(
11.6
), Formulation
mode(
11.7
).
• Cannot be used together with the EP-100/-110 printer statistical
calculation function.