92
14
. Troubleshooting
14.2
Error Display
Error display
Description
Countermeasure
Trouble in weight loading mechanism
Check transportation screws.(Refer to
2.3
)
The load on the pan is unstable at
calibration
Avoid wind and vibration.
The drift of zero point is large at
calibration.
Install the pan properly. Unload the pan.
The drift is large at the time of PCAL
Use correct weight.
The drift is large at span calibration
Use correct weight.
Calibration weight is wrong
Use correct weight.
Failure in the balance
(Stops when this is displayed)
Contact the service company
Received command code is not
correct
Check delimiter etc.(Refer to
12.3
)
Integer of the displayed unit exceeded
7 digits
Decrease the load.
Failure in the balance.
Contact the service company
PCAL password error
Check the password.(Refer to 7.3.5)
Attempted to set improper numeric
value.
Review the numeric value and decimal
places.
Menu item selection was attempted
when menu is locked.
Release menu lock. (Refer to
4.5.3
)
Supply voltage is abnormal.
Check the supply voltage.