44
8
. Environment
Every time the circle in the right part of the [O/T] key is pressed during weighing in
Pouring mode, the environmental setting will be changed and cycles in the
following order. Upon changing the setting, the new setting is briefly displayed as
shown in “[ ]”.
normal environment (“norm E”)
→
unstable environment, (“UnStb E”)
→
very
stable environment, (“StAbL E”)
→
normal environment, (“norm E”).
During use, the current environmental setting of Pouring mode can be confirmed
by observing the position of the symbol in the display.
When
[PrE.AUto]
is selected
When
[UnStb E]
is selected
When
[norm E]
is selected
When
[StAbL E]
is selected
(Display)
Note
While “Pouring” is selected as the Stability and Response setting, taring the
container or zeroing the display cannot be made by pressing the rightmost
part of [O/T] key. Press the center or the left part of [O/T] key for taring or
zeroing.
However, when using animal weighing mode (11.6), the animal weighing mode operation settings and
confirmation display (cond 1 to 3) will take priority.
In addition, this will be used as an indicator to display the status of each mode during animal weighing
mode (11.6), solid specific gravity (11.2), and liquid specific gravity (11.3) weighing, so the environmental
settings for pouring mode will not be displayed.
Standard mode:
Select menu item
24
[EA-Stnd] (Environment/Averaging-STaNDard)
This mode is suitable for weighing in a normal environment. Averaging is fixed
and does not change dynamically as in the [EA-AUto] mode.
Anti-vibration mode:
Select menu item
25
[EA-v ibr] (Environment/Averaging-VIBRation)
Use this mode when the balance is used in a location where there are large
vibrations and the display fluctuates in the [EA-AUto] mode. Response is
deteriorated at small mass amount changes.
Anti-wind mode:
Select menu item
26
[EA-w ind] (Environment/Averaging-WIND)
Use this mode when the balance is used in a location where it is exposed to
airflow that causes the display to fluctuate in the [EA-AUto] mode.
Response deteriorates further than the [EAi-vibr] mode, but weighing is
comparatively stabilized.
Note
If weighing cannot be performed efficiently even with the [EA-wind] mode,
change the installation site of the balance or use the optional windbreak
(large).