![Shimadzu UP1023X Instruction Manual Download Page 98](http://html1.mh-extra.com/html/shimadzu/up1023x/up1023x_instruction-manual_1256959098.webp)
83
12
. Connections and Communications with External Equipment
2
4
1
3
5
6
7
11
13
12
8
9
10
14
(3)
Echo back command
The balance again transmits the character strings of N pieces included between an echo back
command ‘{‘ or ’}’ and the delimiter. An unprocessed echo back command is not left in the receiving
buffer of the balance, N ≤ 30.
Example 5:
{ABCDEFG12345<CR>
…After receiving this command, the balance outputs ABCDEFG12345<CR>.
When using the balance together with a printer, these text strings can be printed to the printer.
(Printing of arbitrary text strings)
Note
Only capital alphabets and a part of symbols (decimal point, symbol etc.)
can be used when printing with an electronic printer. A maximum of 15
characters per line.
(4)
Command codes for Format EB type (menu item
77
) [F-PrEEb]) and Format Old EB type (menu item
78
)
(i)
Commands related to output
D01
Continuous output*
D03
Continuous output with stability information*
D05
Single output
D06
Auto Print setting (type of Auto Print is set separately)
D07
Single output with stability information***
D09
Release of continuous output and Auto Print
(ii)
Commands related to operation keys
POWER
Equivalent to the [POWER] key.
Q
Equivalent to the [POWER] key.
MENU
Equivalent to the [CAL] key.
TARE
Equivalent to the [O/T] key.
T
Equivalent to the [O/T] key.
UNIT
Equivalent to the [UNIT] key.
Equivalent to the [PRINT] key.
POWER+
Equivalent to holding the [POWER] key
MENU+
Equivalent to holding the [MENU] key
UNIT+
Equivalent to holding the [UNIT] key
PRINT+
Equivalent to holding the [PRINT] key
RECALC
Equivalent to the [RECALC] key of the AKB-301 Application Keyboard.
C
Equivalent to the [C] key of the AKB-301 Application Keyboard.
*Not applicable to a verified balance as a legal measuring instrument in using region
**Handshake is “OFF” or “Hardware”
***Using a verified balance as a legal measuring instrument: Only output at stable.
(The balance continuausly outputs
every about 80ms**)