7.2
Power Failures
67
Power Supply Unit
INSTRUCTION MANUAL
7
7.2
Power Failures
When a power interruption occurs, the motor inside the turbo molecular pump immediately
begins regenerative braking. The magnetic bearing will use this generated electricity to keep
functioning and the rotor inside the turbo molecular pump will continue being levitated. The
rotation will slow down due to the regenerative braking and eventually the rotor will be supported
by the touchdown bearing. Table 7-2 shows the number of revolutions and period of time that will
elapse before being supported by the touchdown bearing, when the power fails at the rated
speed.
Table 7-2
Time and Rotational Speed During a Power Interruption Before Being
Supported by Touchdown Bearing
Note 1) Times are typical for regenerative braking from the rated speed.
Actual times will vary depending on vacuum conditions inside the pump and the
rotational speed when the power fails.
Pump Model
Power Supply
Model
Rotational Speed Before
Support by Touchdown
Bearing
Period of Time Before Support
by Touchdown Bearing
Note 1
)
TMP-803M/MC/LM/LMC
TMP-1003M/MC/LM/LMC
EI-D1003M
7200rpm
about 4 minutes
TMP-1103MP/MPC/LMP/LMPC
EI-D1103M
7200rpm
about 4.5 minutes
TMP-1303M/MC/LM/LMC
EI-D1303M
7800rpm
about 5.5 minutes
TMP-2003M/MC/LM/LMC
EI-D2003M
5400rpm
about 10 minutes
TMP-2203M/MC/LM/LMC
EI-D2203M
5400rpm
about 14 minutes
TMP-W2303LMC
EI-D2303M
5400rpm
about 14 minutes
TMP-3203M/MC/LM/LMC
EI-D3203M
5400rpm
about 15.5 minutes
TMP-3403LMC
EI-D3403M
5400rpm
about 17 minutes
TMP-4203LMC
EI-D4203M
4200rpm
about 22 minutes
Summary of Contents for EI-D1003M
Page 9: ...viii 263 13228 This page is intentionally left blank...
Page 19: ...SECTION 1 OUTLINE AND DESCRIPTIONS 6 263 13228 1 This page is intentionally left blank...
Page 20: ...2 2IDENTIFICATION AND FUNCTION 2 1 Power Supply Unit...
Page 23: ...SECTION 2 IDENTIFICATION AND FUNCTION 10 263 13228 2 This page is intentionally left blank...
Page 24: ...3 3CONSTRUCTION AND PRINCIPLE 3 1 Power Supply Unit...
Page 26: ...4 4SPECIFICATIONS 4 1 Power Supply Unit 4 2 Standards Fulfilled...
Page 39: ...SECTION 5 INSTALLATION 26 263 13228 5 This page is intentionally left blank...
Page 77: ...SECTION 6 OPERATION 64 263 13228 6 This page is intentionally left blank...
Page 91: ...SECTION 7 TROUBLESHOOTING 78 263 13228 7 This page is intentionally left blank...
Page 119: ...Appendix A COMMUNICATIONS A 28 263 13228 A This page is intentionally left blank...