SECTION 5
INSTALLATION
18
263-13228
5
5.1
Installation of the Power Supply Unit
5.1.1 Location of the Power Supply Unit
Install and anchor the power supply unit inside a rack, which shall be located at a place where
it is not exposed to direct sun ray and well ventilated. Avoid to locate it at the following places.
(1) Place where it is very humid, dusty and, in addition, oil smoke, vapor, water, etc, are exist.
(2) Place where the power supply unit is exposed to direct sun ray and abnormally high
temperature
(3) Place with high amplitude of vibration and impact
(4) Near chemically active gas and explosive/combustible gas
(5) Place with strong magnetic field and electric field, noisy place, and place with strong
radioactive ray
(6) Unventilatable place
5.1.2 Installation of the Power Supply Unit
When mounting the power supply unit onto the customer's rack, use the front panel installation
holes. The installation panel and screws are to be supplied by the customer.
Prepare the shelf to support the weight of the power supply unit in the rack.
How to install the unit onto a rack
(1) Turn the power supply unit over and remove the four rubber pads.
(2) Pass the power supply unit through the cutout hole in the installation panel and secure it
with four screws (See Fig. 5-2).
(3) In order to ensure that the interior of the power supply unit is cooled sufficiently, leave a gap
inside the rack of at least 30 mm above, 5 mm on either side, and 50 mm to the rear of the
power supply unit (See Fig. 5-1).
(4) The space required for installing the cables is shown in Fig. 5-4.
(5) Reattach the rubber pads if necessary.
Summary of Contents for EI-D1003M
Page 9: ...viii 263 13228 This page is intentionally left blank...
Page 19: ...SECTION 1 OUTLINE AND DESCRIPTIONS 6 263 13228 1 This page is intentionally left blank...
Page 20: ...2 2IDENTIFICATION AND FUNCTION 2 1 Power Supply Unit...
Page 23: ...SECTION 2 IDENTIFICATION AND FUNCTION 10 263 13228 2 This page is intentionally left blank...
Page 24: ...3 3CONSTRUCTION AND PRINCIPLE 3 1 Power Supply Unit...
Page 26: ...4 4SPECIFICATIONS 4 1 Power Supply Unit 4 2 Standards Fulfilled...
Page 39: ...SECTION 5 INSTALLATION 26 263 13228 5 This page is intentionally left blank...
Page 77: ...SECTION 6 OPERATION 64 263 13228 6 This page is intentionally left blank...
Page 91: ...SECTION 7 TROUBLESHOOTING 78 263 13228 7 This page is intentionally left blank...
Page 119: ...Appendix A COMMUNICATIONS A 28 263 13228 A This page is intentionally left blank...