SECTION 6
OPERATION
60
263-13228
6
Fig. 6-9
Arrangement of Remote-Control Connector Pins
(The power supply unit rear panel attachment connectors, as viewed from the front)
6.7.2 Compatibility with Previous Models
6.7.2.1 Replacing the EI-xx03M/MD Power Supply Unit
CAUTION
The transistor output has been discontinued. The pins that were used for that output are now
assigned to different functions.If wires were connected to these pins with previous models, then
simply making the connection as it is could result in a malfunction on the system controller end.
Therefore, please revise the wiring.
To avoid the risk of fire, please remove wiring, if wiring of power supply for the transistor output
is connected to Pin No.1 and 2 of remote-control connector.
Replacing EI-xx03M/MD Power Supply Units As-Is (EI-03 Compatible Mode)
(1) Connectors
The shape and pin count of connectors are mutually compatible and can be utilized as-is.
(2) Pin Configuration
It is compatible with EI-xx03M/MD units, except for the transistor output. (See Table 6-4)
(3) Signal Activity
It is compatible with EI-xx03M/MD units.
When using it, changing into SEMI E74 Standard Mode
Change the "ALARM" signal and "REMOTE" signal activation settings from EI-03 to SEMI E74,
in accordance with the "Remote-Control Signal Settings" instructions included in the (4) Menu
Mode table of Section 6.6 "Software Operation".
(1) Connectors
The shape and pin count of connectors are mutually compatible and can be utilized as-is.
(2) Pin Configuration
The "ALARM" signal pin assignments ("make contact" or "break contact") are the reverse of
EI-xx03M/MD units. If necessary, change the wiring connections.
(3) Signal Activity
The "ALARM" signal and "REMOTE" signal activity differs from EI-xx03M/MD units.
Summary of Contents for EI-D1003M
Page 9: ...viii 263 13228 This page is intentionally left blank...
Page 19: ...SECTION 1 OUTLINE AND DESCRIPTIONS 6 263 13228 1 This page is intentionally left blank...
Page 20: ...2 2IDENTIFICATION AND FUNCTION 2 1 Power Supply Unit...
Page 23: ...SECTION 2 IDENTIFICATION AND FUNCTION 10 263 13228 2 This page is intentionally left blank...
Page 24: ...3 3CONSTRUCTION AND PRINCIPLE 3 1 Power Supply Unit...
Page 26: ...4 4SPECIFICATIONS 4 1 Power Supply Unit 4 2 Standards Fulfilled...
Page 39: ...SECTION 5 INSTALLATION 26 263 13228 5 This page is intentionally left blank...
Page 77: ...SECTION 6 OPERATION 64 263 13228 6 This page is intentionally left blank...
Page 91: ...SECTION 7 TROUBLESHOOTING 78 263 13228 7 This page is intentionally left blank...
Page 119: ...Appendix A COMMUNICATIONS A 28 263 13228 A This page is intentionally left blank...