![Shimadzu EI-D1003M Instruction Manual Download Page 88](http://html1.mh-extra.com/html/shimadzu/ei-d1003m/ei-d1003m_instruction-manual_1256944088.webp)
7.5
Alarm Detection Capabilities
75
Power Supply Unit
INSTRUCTION MANUAL
7
Table 7-6
Table of Alarms
Alarm
Code
LCD Display
Cause
Protective Action
11
11=TD COUNTER LIMIT
Counts of the high speed touch-
down counter exceeded the
specified number.
Start-up impossible
(detected during power
supply self-diagnostics)
12
12=PF COUNTER LIMIT
Counts of the power failure touch-
down counter exceeded the
specified number.
13
13=WRONG TMP MODEL
The combination of the pump and
power supply is wrong.
14
14=AC LOW VOLTAGE
Fall in AC input power voltage.
Regenerative braking
15
15=POWER FAILURE
Power failure.
16
16=TMP:OVERLOAD
After accelerating to 80% of the
designated speed or low-speed
setting, the speed dropped below
80% due to overloading etc.
Deceleration
21
21=TMP TEMP/MB CABLE
Increased pump drive motor
temperature.
MB cable is not connected.
Free run
(motor stop)
22
22=TMP:SENSOR ERROR
Pump rotation signal could not be
detected correctly.
23
23=EI:MOTOR OVERCURR
Overcurrent ran through the motor.
24
24=TMP PUMP TEMP
Pump Temperature
31
31=EI:BR OVERTEMP
Increased temperature inside power
supply unit.
32
32=EI:DC-DC OVERTEMP
Increased temperature inside power
supply unit.
33
33=EI:FAN ERROR
Power supply cooling fan has
stopped.
34
34=EI:INV. OVERCURR
Overcurrent ran through the motor.
35
35=EI:INV. OVERVOLT
Defective circuit in the power
supply.
36
36=EI:DC-DC LOW VOLT
Defective circuit in the power
supply.
Regenerative braking
37
37=EI:DC-DC OVERCURR
Defective circuit in the power
supply.
Deceleration
38
38=EI:DC-DC OVERVOLT
Defective circuit in the power
supply.
43
43=EI:PARAM ERROR
Stored parameters are not correct.
Start-up impossible
(detected during power
supply self-diagnostics)
44
44=EI:CPU ERROR
Error in the CPU for inverter control.
Free run
(motor stop)
45
45=EI:BRAKE OVERTIME
Pump does not stop within the
specified time after the stop
operation.
46
46=MOTOR OVERSPEED
Pump rotation speed is too high.
47
47=EI:R-SPEED ERROR
Pump rotation speed cannot be
detected.
Summary of Contents for EI-D1003M
Page 9: ...viii 263 13228 This page is intentionally left blank...
Page 19: ...SECTION 1 OUTLINE AND DESCRIPTIONS 6 263 13228 1 This page is intentionally left blank...
Page 20: ...2 2IDENTIFICATION AND FUNCTION 2 1 Power Supply Unit...
Page 23: ...SECTION 2 IDENTIFICATION AND FUNCTION 10 263 13228 2 This page is intentionally left blank...
Page 24: ...3 3CONSTRUCTION AND PRINCIPLE 3 1 Power Supply Unit...
Page 26: ...4 4SPECIFICATIONS 4 1 Power Supply Unit 4 2 Standards Fulfilled...
Page 39: ...SECTION 5 INSTALLATION 26 263 13228 5 This page is intentionally left blank...
Page 77: ...SECTION 6 OPERATION 64 263 13228 6 This page is intentionally left blank...
Page 91: ...SECTION 7 TROUBLESHOOTING 78 263 13228 7 This page is intentionally left blank...