- 84 -
7-4. Indicator Test mode
Through this “Test Mode”, you can check basic conditions of Indicator.
This Test consists with total 7 tests.
7-4-1. Enter “Test Mode”
Press
key for 4sec, then display will show “F-Test”.
Under this display, press No.2 key and enter the “Test Mode”.
Under “Test Mode”, please choose each test and check the basic conditions of Indicator.
If you want to exit from each “Test Mode”, press
key.
7-4-2. Test Mode
Test Mode
Contents
Test 1.
Analogue
Value Test
Under “TEST” display, press No.1 key and Enter “TEST1” mode.
Under this mode, you can check the A/D value.
If the A/D value is close to 520,000, or there is no change although
pressing or loading some force on/in weighing part, please check load cell,
load cell, cable, connecter, A/D board.
Test 2.
Key test
Under “TEST” display, press No.2 key and Enter “TEST2” mode.
Press each key, and check the pressed key is operated.
Test 3.
Output Relay Test
Under “TEST” display, press No.3 key and Enter “TEST3” mode.
This Test will be operated automatically from Relay1 to Relay6.
※
This test will operate automatically, so please remove all materials
in/on weighing parts.
If you can not remove materials, please remove relay terminals.