- 22 -
5-3. Simulation Calibration Mode (Calibrate without Test weight)
Through this
“Simulation Calibration Mode”
you can make simple calibration
without Test weight.
This calibration mode uses “Load cells’ max. capacity” and “Max. Output Rate(mV)”, the weight
adjustment degree might be less than “Test weight Calibration”.
The guaranteed resolution of this “Simulation Calibration” is 1/3,000.
Remove
“CAL-BOLT”
on the Front panel and press
“CAL - LOCK S/W”
inside.
Then, you can enter the Calibration Mode with
and press
key to
enter “
Simulation Calibration Mode”
with
and start calibration mode with
pressing
key.
Remove the CAL.Bolt.
Push the CAL. S/W
Press NO.1 key to enter
the “CELL Calibration” mode
Check the “CELL-CAL”. Massage
on display.
※
For the save the each step, press
key, for the cancel or move back, press
key.
Step No.1 - Setting “Capacity of Load cell”.
Enter the Capacity input mode
Check Max. Capacity of Load cell
(Refer the Load cell Lable)
Input the Max. Capacity of Load
cell(same value of Label
Under this step, input Max. Capacity of Load cell. (Not weighing Scale)
The Max. Capacity of load cell is stated on
“Test report”
or
“Label”
.
CELCAL
SeT-CAL