Recommended Action
Type / Cause
Error Message
Error Code
Inspect the network
connection status. If the error
cannot be removed, contact
the FSE.
Warning. The buffer memory for
transmission from the system to
the workstation is full.
The buffer memory of
the controller has
overflowed.
0x9050
0x9051
Contact the network
administrator. Configure the
proper settings
(Automatically obtain /
Manually obtain) for the IP
Address.
Error 3
The IP address was set
to 192.168.200.96
because the instrument
failed to automatically
obtain an IP address.
0x9052
Contact the network
administrator. Configure the
proper settings
(Automatically obtain /
Manually obtain) for the IP
Address.
Error 2.
Renewal/refresh of the
IP address failed.
0x9053
Determine the number of
systems connected in the
system group where the error
occurred.
Warning. Ten or more systems
were connected in the same
group.
The maximum number
of systems (10) that
can be registered in the
same group was
exceeded.
0x9055
Clear the error. If the error
occurs again, contact the FSE.
Error 2. An error occurred in the
communication between the
pump and the controller, and the
communication was
disconnected.
A timeout occurred in
communication
between the pump and
controller.
0x9060
Clear the error.
Warning. An error occurred in the
communication between the
pump and the controller, but the
communication was recovered.
An error occurred in
communication
between the pump and
controller.
0x9061
Clear the error. If the error
occurs again, contact the FSE.
Error 2. An error occurred in the
communication between the
autosampler or column oven and
the controller, and the
communication was
disconnected.
A timeout occurred in
communication
between the
autosampler or column
oven and controller.
0x9062
ExionLC
™
100 System
Hardware User Guide
197 / 219
RUO-IDV-05-1839-C
Error Messages