Table 5-8 Needle Distances Moved
Operation
Direction
Button
The needle moves in 0.1 mm increments.
Left-Right and
Front-Back
The needle moves to the highest point.
Up
First needle drop: 4.0 mm
Second and subsequent drops: 0.2 mm
Down
First needle drop: 4.0 mm
Second and subsequent drops: 1.0 mm
Down (Large)
Note:
Press
Stop
to cancel the adjustment during the procedure. The adjustment is finished without
updating the needle position.
Note: Vertical movement
is for verifying the relative position with the hole by bringing the tip of the
needle closer to a well or a sample vial. This function is not for deciding the needle stroke.
11. Press
Next
The next Position Adjustment screen is shown, and the needle moves near the next position.
12. Repeat step
for fine adjustment of the remaining positions.
ExionLC
™
100 System
Hardware User Guide
153 / 219
RUO-IDV-05-1839-C
Service and Maintenance