Description
Item
Controller panel. Has a status panel for operation using a touch panel, and if the panel is
opened.
1
Status indicator. Shows the operating status of the system using different colors of the light
that it illuminates.
• Green: Analyzable
• Yellow: Preparing for analysis
• Orange: Sleeping
• Blue: During analysis
• Red: Error
1-1
(power button). Turns the power on and off. When the system is powered on, it is
illuminated in white, and when it is off, it is illuminated in orange.
Note:
Even when the system is turned off, a standby current is flowing to the instrument.
For the power supply on the main unit, refer to
.
A log in numeric password is requested on the Start screen. The initial value is "00000". To
change the value, refer to
1-2
Front panel. A semi-transparent panel that enables inspection of an installed sample. If the
front panel is removed, the parts of the autosampler, such as a needle and high-pressure
valve, can be maintained.
2
Sample rack. Holds sample vials and microtiter plates.
2-1
Rack LED. Illuminated if the sample rack is inserted. After the autosampler needle starts to
move, and until injection is completed, the rack LED for injection flashes.
Note:
Do not remove the rack while the rack LED is flashing, because doing so can damage
the needle.
2-2
Pump. When the door is open, it is possible to inspect and maintain the inside of the pump,
such as the degasser, low-pressure gradient valve, and drain valve.
3
Valve mount door. When the door is open, it is possible to inspect and maintain the inside of
the autosampler, such as the measuring pump and low-pressure valve.
4
Column oven door. The door of the column oven. Open this door to install the column.
5
ExionLC
™
100 System
Hardware User Guide
21 / 219
RUO-IDV-05-1839-C
Overview