Figure 4-20 Installing Sample Plates on the Rack
• Vial plate: Install No. 1 in the front left corner.
• Microtiter plate and deep-well plate: Install the A01 well in the front left corner.
Note:
Both plates placed on one sample rack must be of the same type. Placing different types of plates
might cause failure in sample injection.
CAUTION: Potential System Damage. Insert the sample rack fully. If it is improperly
inserted, then the needle might pierce the wrong position and damage the instrument
or become clogged. In addition, the needle might pierce the cap for the sample upon
sample aspiration and cause an error.
3. Insert the sample rack all the way horizontally along the guide.
Note:
Move the sample racks with the microtitre plates installed horizontally, so that the sample does not
spill.
When the sample rack is recognized, the positions of the attached racks (right and left) and the plate types
are shown on the Main Analysis screen.