Recommended Action
Type / Cause
Error Message
Error Code
Clear the error and perform
analysis again.
Error 3
A rack number that is
different from the rack
number set in the
sequence might have
been notified to the
instrument.
0x9078
Inspect the column oven for
an abnormality.
Error 2. During validation and
calibration, the temperature of
the column oven did not stabilize
at the set temperature.
The temperature of the
column oven was not
stable.
0x9079
Install a sample vial for the
sample vial number set in the
sequence.
If
Vial Skip
in the system
settings is set to
Enabled
,
analysis for which no sample
vial has been installed will
not be executed, but the next
analysis will be executed.
Warning. No sample vial is
installed for the sample vial
number set in the sequence.
No sample vial is
installed in the
autosampler rack.
0x907A
Determine whether the
power for the system and the
workstation have been
turned off. If they have not,
verify the power status.
Error 2. The power to the system
was turned off during analysis.
The power to the
instrument was turned
OFF during analysis.
0x907C
Clear the error. If the error
occurs again, contact the FSE.
Warning. No sample vial has been
installed for the sample vial
number set in the next analysis.
There is no sample vial
installed for the next
analysis in the
autosampler rack.
0x907E
Inspect the vacuum pressure
of the degassing unit. If the
vacuum pressure is NG, then
set the method parameter for
the degasser power off and
then on. If the vacuum
pressure is OK, then there is
no problem because the
degasser is automatically
recovered.
Warning
A sealing ability error
occurred in the vacuum
flow path of the
degasser.
0x9100
ExionLC
™
100 System
Hardware User Guide
199 / 219
RUO-IDV-05-1839-C
Error Messages