Figure 4-15 Columns up to 100 mm
2. Connect the columns.
Equilibrate the Column
Adjust the temperature of the column according to the column oven.
1. Set the flow rate to one-fifth the set value on the Main Analysis screen.
Note:
When the column is at a low temperature, starting pumping might cause an error due to pressure
that is too high, or it might damage the column. Pump at a lower flow rate until the oven temperature reaches
the set value.
2. Press
Pump
and
Oven
.
Pump
and
Oven
are illuminated in blue and the operation starts. When the oven temperature reaches the
set value, the status area changes from
Wait Oven
to
Ready
.
3. After the oven temperature reaches the set value, return the flow rate to the set value.