12. Output
12-4. Analog output
12-4-1. Analog output: D/A (standard spec.)
Analog voltage is always output depending on measuring force value. (+/- 2V when
max. force is applied.)
Force value can be recorded at real time by connecting to external equipments with
analog cable (excluded).
Analog output
Data update: 2000 data / sec.
Zero adjustment: /-20mV
Accuracy: 1% or less
* Connect to the external equipments with resistance 1k
Ω
and more.
12-4-2. Analog output: RAW (optional spec.)
The raw analog data is output without digital processing.
The response speed is fast, but zero reset is invalid. (Noise may also be detected as
the data is not filtered.)
Output voltage is approx. +/-1v when max. force is applied. The voltage may differ
depending on load cells of separated sensor model.
* Connect to the external equipments with input resistance 1k
Ω
and more.
* This option is not available with eZT.
* The analog output is unstable when the introduction message shows up
on the display. Please use the analog output during measurement.
12-5. Wireless Transmission Adapter facility
This instrument is compatible with a *Wireless Transmission Adapter (*sold separately)
for the Wireless Transmission of the measurement values.
Settings as follows:
When the Wireless Output ON, there are restrictions such as some functions that
cannot be used at the same time.
*Refer to the Wireless Transmission Adapter Operating Manual for details.
Setup Menu
Wireless Setting
Wireless Output
ON