37
38
12. Output
12-2. USB output (output to PC)
ZTA / ZTS can be connected to PC with included USB cable.
For compatible PCs, please check the operating environment of the software.
12-2-1. Connection to PC
Connect the interface of force gauge and USB port of
PC with the included USB cable.
12-2-2. Installation of driver
Turn on the force gauge while connected to PC.
The force gauge is detected as the new device. Insert the included CD-ROM to PC
and follow the direction of Force Logger Installation manual.
* Installation of driver is necessary for data logger software Force Logger
(included) and graphing software Force-Recorder (optional).
12-2-3. Installation of data logger software Force Logger
Install data logger software Force Logger after installation of driver.
Select CD drive in My Computer and click an icon of Setup. (CD-ROM is still inserted
to PC.) Follow the direction of Force Logger Installation manual to install.
Commands
The command is common among RS232C and USB interface.
The force gauge basically responses after receiving commands.
Commands and responses are consisted of ASCII code.
Commands and responses are followed by code [CR].
The force gauge responses when receive code [CR].
The force gauge sends E[CR] when a wrong command is sent.
Gain with C [CR] code. Please refer to the page 51-58 for commands in
detail.
• Some PC and environment may not correspond to the CD-ROM. Please
get a contact with your local distributor or us in this case.
12-3. Output on RS232C/USB
Connecting with external equipments, data transport and control of force gauges are
possible. The connection is based on RS232C (optional cable) and USB (included
cable).
Data bits
Stop bit
Parity bit
Transmission rate
8 bit
1 bit
None
19200bps
RS232C, Condition