4.8 Configuration of non-weighing point related parameters
The configuration of all parameters, which do not belong to the internal weighing point, is divided into several
sections (see chapter 4.3.1.3 Parameter tree).
4.8.1 Set clock (date and time)
For adjusting date and time, select [Setup]-[Set Clock].
Press
to alter the time, enter numeric values, the colon (:) is
entered with
, confirm with
.
Change to date with
or
. Press
to alter the date, enter
numeric values the dot (.) is entered with
, confirm with
.
Return to the [Setup] with
.
4.8.2 Serial
ports
Configuration of serial interfaces, select [Setup]-[Serial Ports].
[Operator device at]
Switches the operating console to a serial interface connected to
a terminal or PC with terminal program.
Further options depending on the installed interfaces. Select with
or
[Builtin RS232] or [Slot (1 ... 3) RS485 or RS 232] or
[none]; confirm with
.
Error message, if no serial interface is installed in selected slot.
Error message, if serial interface is already assigned to another
device (e.g printer).
If Operator device is selected, interface parameters cannot be changed and are set to:
[Protocol]: XON/XOFF, [Baudrate]: 9600, [Bits]: 8, [Parity]: None, [Stopbits]: 1, [Devtype]: Ansi, [Echo]: Enabled
[Printer device at]
Menu is comparable to [Operator device at]
Select the serial interface connected to the printer.
[Remote device at]
Menu is comparable to [Operator device at]
Select the serial interface connected to the remote device (Example EXTEC terminal TERM300), interface
parameters cannot be changed and are set to:
Raw, [Echo]: Disabled
[Protocol]: XON/XOFF, [Baudrate]: 9600, [Bits]: 8, [Parity]: None, [Stopbits]: 1, [Devtype]: Raw, [Echo]: Disabled