PR 5610 Installation Manual
Operation with Terminal
Sartorius
103
4.10.4 Function 5: Write a bit
Command Device
address
Function
number
Bit address
Bit value
always 0
CRC 16
1 byte
1 byte
2 bytes
1 byte
1 byte
2 bytes
Range
0...127
5
0...32760
0 or 255
0
With device address 0 the command is handled by all connected units, however, no reply is
sent.
Reply Device
address
Function
number
Bit address
Value of
bit
always 0
CRC 16
1 byte
1 byte
2 bytes
1 byte
1 byte
2 bytes
With an address out of the permissible range (0...32760), an error message is sent as a reply.
Example of function 5 for setting bit 140 (taring)
Command 1
5
0
140
255
0
CRC
CRC
Reply 1
5
0
140
255
0
CRC
CRC
The individual bytes are shown.
4.10.5 Function 6: Write a word
Command Device
address
Function
number
Word
address
Value of
word
CRC 16
1 byte
1 byte
2 bytes
2 bytes
2 bytes
Range
0...127 6 0...2047
With an address of 0 the command is handled by all connected units, however, no reply is
sent.
Reply Device
address
Function
number
Word
address
Value of
word
CRC 16
1 byte
1 byte
2 bytes
2 bytes
2 bytes
With an address outside the permissible range (0...2047), an error message is sent as a reply.