5.5 Internal ADC, parameters
The editing functions for parameter entry are described in the first sub-section, the individual parameters are
described in the following sub-sections.
5.5.1 Editing functions for parameter entry
Weighing point selection is done with the cursor keys [
↑
] and [
↓
]. The selected weighing point is highlighted by
inverse display.
The weighing point calibration mask is displayed with different command lines dependent of the CAL-switch
position.
•
Command line: calibration protected
Keys
E
Exit display
Cal - Jumper closed
A note that the CAL-switch is closed is displayed (
Cal-Jumper closed
). I.e. the data can be displayed
without changing them. The mask can be left by pressing key [E] (
E Exit display
).
•
Command line: access permitted
Keys
Select item with cursor
up
/
down
/
left
/
right
Change with
+
/
-
Return to Fi
X
ed,Set
F
ullscale,Calibrate by
W
eight or
M
illivolts
E
Exit calibration
1. The commands for entry of the weighing point parameters are displayed in the first line:
Select item with cursor up/down/left/right Change with +/-
See items Select item and Enter/change parameters' .
2. The second line refers to weighing point calibration.
3. The last line contains the command for leaving the mask
E Exit calibration.
See chapter 5.4.2 'Exit calibration'.
♦
Item selection
Item selection is with the four cursor keys [
←
], [
→
], [
↑
] and [
↓
] (
Select item with cursor up/down/left/right
). The
selected item is displayed inversely.
♦
Parameter entry/ changing
The entry/ change is dependent of the item. The two types of entry/ change are:
•
Selection item: selection of predefined parameters is with keys [+] and [-]
Selection fields are:
Filter, Testmode
and
Stepwidth
. All other fields are normal input items. Nevertheless,
they can be changed with keys [+] and [-].
•
Input item: entry of required parameter value.
During entry, the values are shifted from right to left. Point or comma are part of the mask (cannot be
entered). During entry, the content of the item is underlined. The entry can always be canceled by pressing
the space bar [
_
]. The old value remains unchanged.
The parameter entry is completed with key Enter [
↵
]. Subsequently, the entry is checked:
If an entry is within the permitted range, the value is stored and not underlined any more.
If an entry is not within the permitted range, the entered value is replaced by the next permitted value
and remains underlined.
The operator has the following possibilities: Press key Enter [
↵
] to confirm this next value. Enter a new
value. Redisplay the old value by pressing the space bar [
_
].