Commissioning
PR 5610 Installation Manual
76 Sartorius
Enter the local gravity from 9.00000 to10.00000;
confirm with
.
Select the hysteresis error with
or
[specified] or [not
specified] with
or
; confirm with
. The choice
[specified] is followed by the entry of the correction values, which
is jumped over for [not specified].
Enter the lower correction value A in %; confirm with
.
Enter the upper correction value B in %; confirm with
.
Select the load cell data [all LC same] or [each LC specific] with
or
; confirm with
.
Automatic jump in the entry menu for the load cell data. Enter
the additional load cell data with [Enter].
[Enter]
Enter the sensitivity for load cell 1; confirm with
.
Enter the output resistance for load cell 1; confirm with
. With
[each LC specific] enter these data the same way for all other
load cells.
After that, and also with [all LC same], jump back in the entry
menu for the load cell data.
[Calc]
Calculation of the span value according to the entered data
(Smart-Calculation) with [Calc] .
Display of the span value. Select between confirmation of this
value with [Accept] or jump back in the entry menu for the span
value with [Change].
Acknowledgement after pressing
.
Display of the calculation menu for the test value. Continue with
4.5.11 Calculate test value.
4.5.11 Calculate test value
[Test]
Select [Test] for calculating the test value.
Acknowledgement after [Test], weight display shows FSD.