Commissioning with Terminal
PR 5610 Installation Manual
130 Sartorius
•
Set the fixed span
X to set fixed span
If the calibration cannot be finished correctly for any reason, press key [X]. To avoid operator errors, a
prompt is displayed and must be replied with Yes/ No:
[Y/N]
.
The system executes the command. Subsequently, the command line as shown in point 5 is displayed
again.
•
Print the actual weight:
P to Print
Press key [P]. The system executes the command. Subsequently, the command line as shown in point 5
is displayed again.
•
Set
the
zero
Unload and hit O
If necessary, zero setting (as described in point 2.) can be repeated.
•
Finish
calibration
E to End
Press key [E]
Calculating test …
the system calculates the test value
Calculating test … Status OK
Subsequently, the standard command line is displayed:
Keys
Select item with cursor
up
/
down
/
left
/
right C
hange with
+
/
-
Return to Fi
X
ed, Set
F
ullscale, Calibrate by
W
eight or
M
illivolts
E
Exit calibration
Calibration is finished. The data must be saved: Proceed as described in points 9. and 10. of the 'Weighing point
calibration procedure'.
5.6.6.2 Subsequent deadload correction
If the hopper weight changes e.g. due to loss (deadload reduction) or slag (deadload increase) by an amount
which is higher than the zero set range, the automatic zero tracking and manual zero setting will not function
any more. In these cases, subsequent deadload correction is possible, without recalibrating.
For this, call up the weighing point calibration and set only the deadload. Calibration must be skipped.
Zero correction
Proceed as described in points 1. to 7. of the 'Weighing point calibration procedure'. This is followed by actual
calibration: zero correction:
1. The following command line is displayed in the calibration mask:
Unload
Calibration by weight
Unload and hit
O
or
S
to skip,
Q
to quit,
E
to end,
X
to set fixed span,
P
to print
The system offers the following replies:
Unload and hit 0
unload scale, set the zero
S to skip
skip zero setting
Q to quit
leave the calibration without doing anything
E to end
calibrate and finish
X to set fixed span
set the fixed span
P to Print
print out the weight
2. Set the deadload
•
1. Unload the scale
•
2. Press key [0]
•
The instrument evaluates the data. The following command line is displayed in the calibration mask:
Setting deadloads …
s
ystem saves the data, subsequently:
Setting deadloads … Status OK
message is displayed during approx. 2 sec. The weight display goes to:
0000 kg
•
In case of error, repeat zero setting. See point 3