
Hardware Configuration
PR 5610 Installation Manual
22 Sartorius
3.3.1 Load cell cable / cable junction box
Connection of up to 8 load cells (650 Ohm) in 6-wire technique:
•
via cable junction box PR6130/.. with cable PR6135/..
The here mentioned
cable colours are valid for
SARTORIUS load cells of series
PR 62 . . resp. for extension
cable PR 6135.
For use of Sartorius platforms
or other load cells the cable
colours could have a different
meaning!
Before connecting please refer
to the respective manual /
datasheet for the meaning of
the cable colours.
Recommendation:
- lay in steel tubes connected to earth potential.
- min. 1 m distance to power cables
Load cell supply circuit
Load cell circuit resistance load > 75 Ohm e.g. 8 load cells, each with 650 Ohm,
firmly adjusted to 12 V, protected by a 250 mA slow blowing fuse (see chapter 1.3.6).