PR 5610 Installation Manual
Commissioning with Terminal
Sartorius
127
5.6.2 Editing functions for parameter entry
♦
Weighing point selection
Weighing point selection has to be done before calling up this weighing point calibration mask: Weighing point
selection is by means of the two cursor keys [
↑
] and [
↓
]. Thereby, the actually selected weighing point is
highlighted by inverse display.
♦
Item selection
Direct access is only to item
Fullscale
. Item selection is by entry of a letter (unlike the cursor keys as for
weighing point configuration parameters). Parameters
Resolution
are not accessible (only used for informative
purposes).
♦
Entry/ change of parameter
•
Enter the required parameter value: value and + dimension
•
Change the entry by overwriting (select character with the cursor keys). Delete by pressing the
backspace key [
←
] and re-enter.
•
Press the enter key [
↵
] to complete the entry.
•
Subsequently, the entry is checked.
5.6.3 Adjusting the full scale value:
Fullscale
Enter the full scale value into item
Fullscale
.
♦
Open the entry dialogue by entry of letter 'F'.
♦
Prompt for entry of the measurement value:
Enter new fullscale [3000kg ], E to exit
: The old measuring
range with dimension is displayed [in square brackets] (default value:
3000 kg
). the instrument expects the
following entries:
- Measuring range with comma (or point)
- no or one or several spaces
- dimension (g, kg, lb or t)
Complete by pressing the enter key [
↵
].
♦
The instrument checks the entry for plausibility. If the entered full scale is lower than the calibration weight
specified during calibration (
Calipoint
) or if the calculated input voltage for the specified full scale is out of
the permissible limits, the new full scale value is rejected. The full scale is rejected, if less than 0.8 counts/d
are provided.
•
If the entry is accepted, the following message is displayed during approx. 1 sec:
Status OK
•
Error messages:
Bad Weight
e.g. dimension missing
Status Arithmetik overflow
e.g. weight value too high
Status Above phymax
calculated input voltage: > 36 mV
Too many d
resolution too low: < 0.8 c/d
Status Below Cal
value below calibration point
Illegal Fullscale
the last two digits are unequal to 00 or have more than 3
digits behind the decimal point
♦
Leave input item without changing the old measurement value (with operator error):
•
If necessary, delete entries: position the cursor in the leftmost position with the cursor key [
←
].
Delete all entries with the delete key [Del].
•
Enter key [
↵
]: The old span remains unchanged.
⇒
Adjusting the stepwidth (divisions):
Stepwidth