RTR-4 Operator’s Manual
System Operation
120300 Rev. D
SAIC Proprietary
4-5
• The RTR-4 imaging software has been properly configured and
is initiated and running.
Procedure RTR0006
Obtain a Added image as follows:
STEP
ACTION
1
Close all images currently open.
2
Click on the
Preferences
pull-down menu.
3
Verify the
Multiple Originals
command is disabled
(unchecked.)
WARNING
Excessive exposure to X-Ray radiation is harmful.
Verify the exclusion zone is clear of all personnel.
Failure to comply may result in personnel injury.
4
Acquire the first image and save it.
CAUTION
If the source is set to fire more than 99 pulses, wait
five minutes before firing the source again. This
permits the source to cool properly and allow the
battery to recover. Failure to comply may damage
the source unit.
RTR0006a